CN-0326 Software User Guide
OVERVIEW
The circuit is a completely isolated low power pH sensor signal conditioner and digitizer with automatic temperature compensation for high accuracy. The circuit gives 0.5% accurate pH readings for pH values from 0 pH to 14 pH with greater than 14-bits of noise-free code resolution and is suitable for a variety of industrial applications such as chemical, food processing, water, and waste water analysis. This circuit supports a wide variety of pH sensors that have very high internal resistance that can range from 1 MΩ to several TGΩ. , and digital signal and power isolation provides immunity to noise and transient voltages often encountered in harsh industrial environments.
GENERAL SETUP
- EVAL-CN0326-PMDZ Evaluation Board
- EVAL-SDP-CB1Z Evaluation Board
- SDP-PMD-IB1Z Evaluation Board
-
- (supplied with provided CD in kit)
- +6V DC power supply
- USB Type-A plug to USB Mini-B plug cable
- pH probe with BNC connector
- RTD sensor with RCA connector
MINIMUM PC/SYSTEM REQUIREMENTS
- One PC with the following
HOW TO INSTALL THE EVALUATION SOFTWARE
- Extract the file CN0326 Eval Software.zip and open the file setup.exe.
NOTE: It is recommended that you install the CN0326 Evaluation Software to the default directory path C:/Program Files/Analog Devices/CN0326/ and all National Instruments products to C:/Program Files/National Instruments/
- Upon completion of the installation of the CN0326 Evaluation Software, the installer for the ADI SDP Drivers will execute.
NOTE: It is recommended that you close all other applications before clicking “Next”. This will make it possible to update relevant system files without having to reboot your computer.
STEP BY STEP INSTRUCTION FOR CONNECTING THE HARDWARE
- Plug the mini end of the USB cable into connector J1 of the EVAL-SDP-CB1Z and connect the other end of the USB cable into the Laptop or PC.
- Remove the jumper across JP1 found on the SDP-PMD-IB1Z.
- Connect the 120-pin connector of the SDP-PMD-IB1Z circuit board to the connector marked “CON A” on the EVAL-SDP-CB1Z evaluation (SDP) board. Nylon hardware should be used to firmly secure the two boards, using the holes provided at the ends of the 120-pin connectors.
- Connect the EVAL-CFTL-6V-PWRZ to J1 barrel jack of the SDP-PMD-IB1Z, and WAIT 10 SECONDS before continuing
- Connect the 12-pin right angle male pin header of EVAL-CN0326-PMDZ to the 12-pin right angle female pin header of SDP-PMD-IB1Z.
- Connect the pH probe in J1 of EVAL-CN0326-PMDZ.
- Connect the RTD sensor in P1 of EVAL-CN0326-PMDZ.
- Place the jumper across JP1 on the SDP-PMD-IB1Z and make sure it's positioned to +3.3V.
OPENING AND ENABLING THE EVALUATION SOFTWARE
- Launch the executable found at C:/Program Files/Analog Devices/CN0326 and press the “Connect” button.
- After pressing the “Connect” button, a prompt will appear informing the user that the SDP is now ready to acquire data and the rest of the buttons found in System Controls will be enabled.
- Press the “Run” button in order to start using the program.
STEP BY STEP INSTRUCTION USING THE EVALUATION SOFTWARE
- Below are the list of available software controls grouped according to their location in the software GUI:
- System Controls
- Run – This button polls the data ready bit in the AD7793 status register. When this bit is set, LabVIEW reads the data registers and displays the pH reading and the temperature data on the chart found in the evaluation software.
- Save Data – This button saves all data displayed in the LabVIEW chart.
- Clear Data – This button clears all data displayed in the LabVIEW chart.
- Data Graph – This page contains a waveform chart visually displaying the pH readings and the temperature readings taken after clicking the Run button.
- Calibration Settings – This page contains the controls necessary to calibrate the entire pH measurement system using a two-point linear interpolation.
- Two Point Calibration – This button will calculate the calibration coefficients based on two data points gathered on the Sampled Data page.
- Voltage Offset Calibration – Removes the offset introduce by the pH probe and by the System after clicking the Remove Offset Voltage button.
- Buffer Solution 1 – This gives the customer an option to choose the source of the list of table of the temperature coefficient for buffer solution 1 that will be used for calibration. This option also set the first point for calibration after clicking the Calibration Point 1.
- NIST Standard Buffer Solution – allows the customer to choose the predefined list of table for the buffer solution traceable by the NIST.
- Custom Buffer Solution - allows the user an option to choose a table that gives the temperature coefficient of the buffer solution to be used for calibration.
- Buffer Solution 2 – This gives the customer an option to choose the source of the list of table of the temperature coefficient for buffer solution 2 that will be used for calibration. This option also set the second point for calibration after clicking the Calibration Point 2.
- NIST Standard Buffer Solution – allows the customer to choose the predefined list of table for the buffer solution traceable by the NIST.
- Custom Buffer Solution - allows the user an option to choose a table that gives the temperature coefficient of the buffer solution to be used for calibration.
- Measured Results –Display the result for pH reading, the corresponding voltage for the pH reading and the slope expressed in mV/pH.
- Continuous Temperature Compensation – This allows the user to enable or disable continuous temperature compensation for pH reading.
- Data Rate – This allows the user to modify the rate of data acquisition which changes instantly by clicking Refresh Rate button.
- RTD Value – This allows the user to use other RTD value according to their requirements.
- Save/Recall Calibration Settings – this allows the user to save setup by clicking the Save button or recall the calibration setup formatted in a text file.
- To save the calibration settings, click Save button then a choose the desired location and provide a the file with a text file(.txt) extension to create new file or click to over write an existing file with a .txt extension.
- To recall the calibration setting, click the folder to browse for the location of the text file that contains the calibration settings and click Load to change the settings indicated in the file that was loaded.
- SDP Board Firmware Revision – provides details on the firmware version of the Blackfin used by the SDP board.
- Note: Example of a format for the calibration setup can be found in the Design Support Package with a folder name Calibration Text Format
CALIBRATION
Pre-Calibration Note:
- A minimum of two buffer solution of different pH values have to be selected.
- One of these buffer solutions should be pH 7 for the zero point compensation.
- The second buffer solution should have a pH value which lies as near as possible to the anticipated measuring range.
- The third buffer solution will be needed if the expected measuring range covers the acidity and alkalinity range where in the pH value of the second and third buffer solution should fall under the acid region and alkaline region respectively.
- The value of the chosen buffer solution should at least differ by 2 pH units.
- The electrode assembly of a pH probe has to stay in the buffer solution for a short time till the indication of the pH meter become stable.
This section explains the steps required to properly calibrate the EVAL-CN0326-PMDZ pH measurement system.
- Step 1
- Immersed the pH probe along with the temperature sensor into the buffer solution with pH value of 7 for zero point compensation.
- Click Remove Offset Voltage button once the pH voltage reading becomes stable.
- Step 2
- After rinsing the electrode assembly with deionized or distilled water, immersed the pH probe with the temperature sensor into the second chosen buffer solution.
- Select the source of the buffer solution temperature coefficient table for compensation.
- When choosing the predefined table preprogrammed in this software, choose the pH value that corresponds to the buffer solution that will be used to define the second point.
- When choosing the custom buffer solution table, navigate through the panel that contains the file that will be used for temperature compensation. (Note: a sample format of a table can be found in CN0326 Design Support Package with a folder name NIST Buffer Solution Table Format)
- After choosing the source of the buffer solution temperature coefficient table, click Calibration Point 1 button to set the second point of calibration.
- Step 3
- After rinsing the electrode assembly with deionized or distilled water, immersed the pH probe with the temperature sensor into the third chosen buffer solution.
- Select the source of the buffer solution temperature coefficient table for compensation.
- When choosing the predefined table preprogrammed in this software, choose the pH value that corresponds to the buffer solution that will be used to define the third point.
- When choosing the custom buffer solution table, navigate through the panel that contains the file that will be used for temperature compensation. (Note: a sample format of a table can be found in CN0326 Design Support Package with a folder name NIST Buffer Solution Table Format)
- After choosing the source of the buffer solution temperature coefficient table, click Calibration Point 2 button to set the third point of calibration.
- CN0335軟件用戶指南
- CN0336軟件用戶指南
- CN0349軟件用戶指南
- CN0350軟件用戶指南
- CN-0354軟件用戶指南
- CN0355軟件用戶指南
- CN0196軟件用戶指南
- CN0203軟件用戶指南
- CN0204軟件用戶指南
- CN-0269軟件用戶指南
- CN-0271軟件用戶指南
- CN-0288軟件用戶指南
- CN-0296軟件用戶指南
- CN-0301軟件用戶指南
- 基于CN-0326酸鹼電極的參考設計
- 上位機軟件的用戶界面設計技巧 773次閱讀
- 搞不懂組態軟件和HMI?常見組態軟件有哪些? 4789次閱讀
- 常用的遠程控制軟件介紹 1763次閱讀
- 醫療器械軟件風險分析介紹 醫療設備軟件靜態和動態分析技巧 1850次閱讀
- 如何使用openEuler用戶軟件倉(EUR) 1223次閱讀
- 基于STM32 MCU應用的EMC指南之軟件篇 1020次閱讀
- 開源、可重用的軟件堆棧支持CbM的實時處理和算法開發 900次閱讀
- EsDA為用戶帶來簡單易用的軟件設計體驗 1884次閱讀
- 使用含快速開關SiC器件的RC緩沖電路實用解決方案和指南 2531次閱讀
- 常見的組態軟件有哪些 1.2w次閱讀
- 如何正確的設置IAR單片機編程軟件 4994次閱讀
- 一款以BitTorrent代幣連接和獎勵用戶的軟件BitTorrent Speed介紹 3328次閱讀
- 組態軟件的特點和應用 1.4w次閱讀
- 組態軟件是什么_組態軟件有什么用 3.5w次閱讀
- 長虹CN-7機芯彩電場掃描電路分析 1w次閱讀
下載排行
本周
- 1電子電路原理第七版PDF電子教材免費下載
- 0.00 MB | 1491次下載 | 免費
- 2單片機典型實例介紹
- 18.19 MB | 95次下載 | 1 積分
- 3S7-200PLC編程實例詳細資料
- 1.17 MB | 27次下載 | 1 積分
- 4筆記本電腦主板的元件識別和講解說明
- 4.28 MB | 18次下載 | 4 積分
- 5開關電源原理及各功能電路詳解
- 0.38 MB | 11次下載 | 免費
- 6100W短波放大電路圖
- 0.05 MB | 4次下載 | 3 積分
- 7基于單片機和 SG3525的程控開關電源設計
- 0.23 MB | 4次下載 | 免費
- 8基于AT89C2051/4051單片機編程器的實驗
- 0.11 MB | 4次下載 | 免費
本月
- 1OrCAD10.5下載OrCAD10.5中文版軟件
- 0.00 MB | 234313次下載 | 免費
- 2PADS 9.0 2009最新版 -下載
- 0.00 MB | 66304次下載 | 免費
- 3protel99下載protel99軟件下載(中文版)
- 0.00 MB | 51209次下載 | 免費
- 4LabView 8.0 專業版下載 (3CD完整版)
- 0.00 MB | 51043次下載 | 免費
- 5555集成電路應用800例(新編版)
- 0.00 MB | 33562次下載 | 免費
- 6接口電路圖大全
- 未知 | 30320次下載 | 免費
- 7Multisim 10下載Multisim 10 中文版
- 0.00 MB | 28588次下載 | 免費
- 8開關電源設計實例指南
- 未知 | 21539次下載 | 免費
總榜
- 1matlab軟件下載入口
- 未知 | 935053次下載 | 免費
- 2protel99se軟件下載(可英文版轉中文版)
- 78.1 MB | 537793次下載 | 免費
- 3MATLAB 7.1 下載 (含軟件介紹)
- 未知 | 420026次下載 | 免費
- 4OrCAD10.5下載OrCAD10.5中文版軟件
- 0.00 MB | 234313次下載 | 免費
- 5Altium DXP2002下載入口
- 未知 | 233046次下載 | 免費
- 6電路仿真軟件multisim 10.0免費下載
- 340992 | 191183次下載 | 免費
- 7十天學會AVR單片機與C語言視頻教程 下載
- 158M | 183277次下載 | 免費
- 8proe5.0野火版下載(中文版免費下載)
- 未知 | 138039次下載 | 免費
評論
查看更多